BD

Becton, Dickinson and Company (BD)

CVEs, 365 days022 all time, 0 rejected
Quality-median 99%
CVSS by the CNA-share of records
CWE-share of records
Affected versions-not n/a or unspecified
CVSS added by CISA-lower is better
OrganisationBecton, Dickinson and Company (BD)
RoleCNA
TypeVendor
Reports toicscert → CISA
ScopeBD software-enabled medical devices only.
Advisorieshttps://cybersecurity.bd.com/bulletins-and-patches
  • CVE-2024-10476 2024-12-17 · CVSS 8.0
    Default credentials are used in the above listed BD Diagnostic Solutions products. If exploited, threat actors may be able to access, modify or delete data, inc
  • CVE-2023-29066 2023-11-28 · CVSS 3.2
    The FACSChorus software does not properly assign data access privileges for operating system user accounts. A non-administrative OS account can modify informati
  • CVE-2023-29065 2023-11-28 · CVSS 4.1
    The FACSChorus software database can be accessed directly with the privileges of the currently logged-in user. A threat actor with physical access could potenti
  • CVE-2023-29064 2023-11-28 · CVSS 4.1
    The FACSChorus software contains sensitive information stored in plaintext. A threat actor could gain hardcoded secrets used by the application, which include t
  • CVE-2023-29063 2023-11-28 · CVSS 2.4
    The FACSChorus workstation does not prevent physical access to its PCI express (PCIe) slots, which could allow a threat actor to insert a PCI card designed for
  • CVE-2023-29062 2023-11-28 · CVSS 3.8
    The Operating System hosting the FACSChorus application is configured to allow transmission of hashed user credentials upon user action without adequately valid
  • CVE-2023-29061 2023-11-28 · CVSS 5.2
    There is no BIOS password on the FACSChorus workstation. A threat actor with physical access to the workstation can potentially exploit this vulnerability to ac
  • CVE-2023-29060 2023-11-28 · CVSS 5.4
    The FACSChorus workstation operating system does not restrict what devices can interact with its USB ports. If exploited, a threat actor with physical access to
  • CVE-2023-30565 2023-07-13 · CVSS 3.5
    An insecure connection between Systems Manager and CQI Reporter application could expose infusion data to an attacker.
  • CVE-2023-30564 2023-07-13 · CVSS 6.9
    Alaris Systems Manager does not perform input validation during the Device Import Function.
  • CVE-2023-30563 2023-07-13 · CVSS 8.2
    A malicious file could be uploaded into a System Manager User Import Function resulting in a hijacked session.
  • CVE-2023-30562 2023-07-13 · CVSS 6.7
    A GRE dataset file within Systems Manager can be tampered with and distributed to PCUs.
  • CVE-2023-30561 2023-07-13 · CVSS 6.1
    The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially
  • CVE-2023-30560 2023-07-13 · CVSS 6.8
    The configuration from the PCU can be modified without authentication using physical connection to the PCU.
  • CVE-2023-30559 2023-07-13 · CVSS 5.2
    The firmware update package for the wireless card is not properly signed and can be modified.
  • CVE-2022-47376 2023-06-13
    The Alaris Infusion Central software, versions 1.1 to 1.3.2, may contain a recoverable password after the installation. No patient health data is stored in the
  • CVE-2022-43557 2022-12-05 · CVSS 5.3
    The BD BodyGuard™ infusion pumps specified allow for access through the RS-232 (serial) port interface. If exploited, threat actors with physical access, specia
  • CVE-2022-40263 2022-11-04 · CVSS 6.6
    BD Totalys MultiProcessor, versions 1.70 and earlier, contain hardcoded credentials. If exploited, threat actors may be able to access, modify or delete sensiti
  • CVE-2022-30277 2022-06-01 · CVSS 5.7
    BD Synapsys™, versions 4.20, 4.20 SR1, and 4.30, contain an insufficient session expiration vulnerability. If exploited, threat actors may be able to access, mo
  • CVE-2022-22767 2022-06-01 · CVSS 8.8
    Specific BD Pyxis™ products were installed with default credentials and may presently still operate with these credentials. There may be scenarios where BD Pyxi
  • CVE-2022-22765 2022-02-12 · CVSS 8.0
    BD Viper LT system, versions 2.0 and later, contains hardcoded credentials. If exploited, threat actors may be able to access, modify or delete sensitive inform
  • CVE-2022-22766 2022-02-11 · CVSS 7.0
    Hardcoded credentials are used in specific BD Pyxis products. If exploited, threat actors may be able to gain access to the underlying file system and could pot

All CVE Program partners